ASTM D5453 Ultraviolet Fluorescence Sulfur-in-Oil content Analyzer

Detailed product descriptionStandard:SH/T0689. ASTM D5453. Etc.Measurement Range:1mg/L~10000 Mg

WhatsApp : +86 13816217984

Email : info@qinsun-lab.com

İçerik

Detailed product description

Standard:SH/T0689. ASTM D5453. Etc.Measurement Range:1mg/L~10000 Mg/L (more Than 5000 Mg/L Dilution Method)
Measurement Accuracy:≤ 15% At 1.0 Mg/L~10.0 Mg/L >≤ 10% At 10 Mg/LTemperature Control Range:Room Temperature~1100 ℃
Name:Ultraviolet Fluorescence Sulfur-in-Oil Content AnalyzeKeywords:Oil Analysis Equipment
Brand:STDPower Supply Voltage:220V ± 5%/50HZ.
Power Supply Current:15AHigh Purity Oxygen:Purity ≥ 99.999%
High Purity Argon:Purity ≥ 99.999%
Highlight:

oil analysis equipment

oil testing equipment

Product description

Ultraviolet fluorescence sulfur analyzer is a new generation of precision experimental instrument developed based on the principle of ultraviolet fluorescence and computer technology. With wide linear range, strong anti-interference ability, simple operation, stable and reliable results and other outstanding advantages, it can be widely used in petroleum, chemical industry, electric power, coal, food, environmental protection and other fields. It is an advanced sulfur analyzer at home and abroad.

Applicable standards

SH/T0689. ASTM D5453. etc.

Technical features

To ensure the normal operation of the instrument, the following environmental conditions are required:

1. The ambient temperature is basically constant and cannot be higher than 30 ℃.

2. The ambient air shall not contain corrosive gas and a large amount of dust to avoid component corrosion and short circuit.

3. The working table of the instrument shall be spacious, well ventilated and free from direct sunlight.

Power requirements

1. Power supply voltage 220V ± 5%/50HZ.

2. The power supply current is greater than 15A.

3. It shall have a good ground wire and the resistance to ground shall not be greater than 3 Ω.

Air source requirements

1. High purity oxygen: purity ≥ 99.999%

2. High purity argon: purity ≥ 99.999%

Technical parameters

1. Measurement range: 1mg/L~10000 mg/L (more than 5000 mg/L dilution method)

2. Measurement accuracy:

≤ 15% at 1.0 mg/L~10.0 mg/L

>≤ 10% at 10 mg/L

Temperature control range: room temperature~1100 ℃, independent temperature control system, no water cooling

Sorumluluktan vazgeçme: Yukarıdaki içerik yalnızca endüstri içindeki kişiler arasında referans ve iletişim amacıdır ve doğruluğunu veya eksiksizliğini garanti etmez. İlgili yasa ve düzenlemelere ve bu web sitesinin düzenlemelerine göre, ilgili ürünleri satın alan birimler veya bireyler geçerli nitelikler ve nitelik koşulları almalıdır.