Features

operating pressure (up to 140 bar), resp. flow volume (up to 15 l/min)

test medium temperature (up to 70. resp. 80°C)

nozzle

nozzle distance

spray angle

spray duration

standards

DIN EN ISO 16925:20

DBL 5416

Ford BO 160-04

GM 9531

Volvo STD 423-0015

VW PV 1503

BMWAA-0136

Renault D 252018

PSA D 255376

MAN M 3022

Technical parameters

All parts of the steam jet test chamber in contact with the medium are rust-proof and can be operated with deionized water.

Pump body:

Pump - with direct drive motor and ceramic plunger, high safety, low wear, safety valve, including control motor protection injection time can be set by a timer in the display, lasting from 1 second to 9 minutes

The pressure can be adjusted externally via a pressure control valve. Pressure depends on the spray - see manufacturer's specifications

Technical data: Flow rate :14.8 L/min

Working pressure P:30-140bar can be adjusted by external pressure regulator

Motor N:6.5 kW, 400V Plug :32 amps, 5 poles

Water supply :2-3 bar, 15°C

Nozzle distance :0-450mm

Test Angle :90° -45 ° -30 °+/ -2 °, door with inspection glass, exhaust bore 102mm, fitting, discharge port with dirt filter.

Bottom frame: stainless steel base with housing and 4 guides.

Sterilization water change program: Execute the water change program every 7 days, automatic cycle sterilization water change program - through manual water change, the water in the tank is automatically heated to 68 ° C, when the tank is heated to 68 ° C temperature water is sprayed out through the nozzle. Once the tank reaches 70 ° C, the automatic cycle of the sterilization and water change procedure will be stopped, and after completion, the tank must be cooled to the operating temperature.

Door lock device: The door is automatically electromagnetic locked during pump operation. The door cannot be opened while the pump is running

Heating container: heating power 9.0 kW, maximum temperature 80°C +/-2. digital display,

preheating time :1 hour

Sorumluluktan vazgeçme: Yukarıdaki içerik yalnızca endüstri içindeki kişiler arasında referans ve iletişim amacıdır ve doğruluğunu veya eksiksizliğini garanti etmez. İlgili yasa ve düzenlemelere ve bu web sitesinin düzenlemelerine göre, ilgili ürünleri satın alan birimler veya bireyler geçerli nitelikler ve nitelik koşulları almalıdır.