Mask Synthetic Blood Penetration Tester - Qinsun
Professional Testing Instruments Manufacturer
Ulusal danışmanlık hattı:+86 021-64208466

Mask Synthetic Blood Penetration Tester

Medical Face Mask Synthetic Blood Penetration Resistance TesterMedical masks synthetic blood penetra

WhatsApp: +86 13816217984

E-posta adresi:info@qinsun-lab.com

İçerik
Medical Face Mask Synthetic Blood Penetration Resistance Tester

Medical masks synthetic blood penetration tester is used to determine the resistance of medical protective masks to penetration by synthetic blood at different test pressures,and can also be used to determine the resistance to blood penetration of other coating materials.


Application
 
Mask synthetic blood penetration tester is used for anti-synthetic blood penetration of medical surgical masks, and mainly used in the field of medical protection products. It can also be used for other textile coating material determination.


Blood penetration resistance tester Standards:

GB 19083-2010 Technical requirements of medical face masks - Anti-penetration property of 5.5 synthetic blood.
YY/T 0691-2008 Clothing for protection against infectious agents-Medical face masks - Test method for resistance against penetration by synthetic blood (fixed volume, horizontally projected)  
YY 0469-2011 Technical requirements of medical face masks– Blood penetration test apparatus.
ISO 226092004 Clothing for protection against infectious agents - Medical face masks - Test method for resistance against penetration by synthetic blood (fixed volume, horizontally projected)  
ASTM F1862-07Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Know Velocity)


parameter:

ItemsParameters
Injection distance300mm~305mm, adjustable
Nozzle diameter0.84mm
Jet velocity450cm/s,550cm/s,635cm/s
 

1.The protruding sample fixture can simulate the actual use condition of the mask. Reserve the test target area without destroying the sample, and   distribute synthetic blood in the sample target area.
2.A special jet device with constant pressure can spray a certain volume of synthetic blood in a controlled time.
3.It can fully simulate the corresponding injection rate to test under the average blood pressure of the human body at 10.6kPa, 16kPa, 21.3kPa.
4.The fixed target plate can block the high pressure part of the injected liquid flow to allow the steady flow to be sprayed to the sample, increasing the accuracy and repeatability of liquid spraying speed.

Sorumluluktan vazgeçme: Yukarıdaki içerik yalnızca endüstri içindeki kişiler arasında referans ve iletişim amacıdır ve doğruluğunu veya eksiksizliğini garanti etmez. İlgili yasa ve düzenlemelere ve bu web sitesinin düzenlemelerine göre, ilgili ürünleri satın alan birimler veya bireyler geçerli nitelikler ve nitelik koşulları almalıdır.